1. Product Overview
- Product Name: SM-PM09 Multi-parameter Patient Monitor (15 inches)
- Technical Data Sheet Version: Release V2.4
- Applicable Environments: Adult, pediatric, and neonatal intensive care; anesthesia and peri-operative care; OR and cardiac care environments
- Certifications: CE/ISO/RoHS/CFDA
2. Core Features
2.1 Core System
- Newly advanced A9 main board with Linux OS, calculating speed is 4 times faster than traditional products
- Support storage of 2160 hours trend table and graph review, 2 hours waveform review, 2000 groups NIBP review and 2000 alarm events review
2.2 Body & Display
- 15 inch high-brightness TFT LED
- Support display 9~13 waveforms
- Support 7 channel ECG waveform display simultaneously
- Optional HDMI output
2.3 Printer (Built-in)
- Type: High-speed 50mm thermal printer
- Brand: SEIKO (Japan)
2.4 Central System
- Optional Build-in wireless network module, supporting wired or wireless connection to the central monitoring station
- Optional Support HL7 (Health Level Seven)
2.5 Alarm
- Double high light alarm light bar
- Three-level acousto-optic alarm
- Sensor-off alarm
- Paper out alarm
- Support alarm review
- Support alarm pause
2.6 Linux OS Functions
- Support operation with mouse and computer keyboard (option)
- Multi-display mode to meet different clinical requirement, including standard interface, big font, OxyCRG, trend graph, NIBP review and full leads ECG interface
- NIBP self-test mode: including overpressure test, static pressure test and air leakage test
- Generate ID automatically when register a new patient
- Support medical history search by patient ID, name and mobile number
- 18 types of Arrhythmia analysis and real-time S-T segment analysis and pacemaker detection
- Drug calculation and titration table
- Support multi-language display
- Support online software upgrading by net/USB
2.7 Service Features
- The Support Tool helps technical personnel to – carry out configuration, upgrades and troubleshooting on an individual monitor; back up the monitor settings
- The Service Mode is password-protected and ensures that only trained staff can access service tests and tasks. It includes Skin Type, Brightness Setting, Clear Data and NIBP Calibration
- The Support Tool uses the USB interface of the monitor for software upgrading
3. Performance Specifications
3.1 Dimension and Weight
- Dimension: 393mm382mm173mm
- Weight: 4.5kg (excluding accessories)
3.2 Power Supply
- Voltage: AC100~240V, 50/60HZ
- Power: ≤60W
3.3 Display
- Screen: 15″ color TFT LED
- Resolution: 1024*768 pixels
3.4 Battery (Pluggable)
- Type: Rechargeable lithium battery 14.8V/2200mAh
- Charge Cycle: ≥500 times
- Working time: 2 hours (optional larger battery for 4-5 hours)
3.5 Recorder (Option)
- Method: Thermal printer
- Paper width: 50 mm (1.97 in)
- Printing speed: 12.5/ 25/ 50 mm/s
- Trace: Max. 3 tracks
- Recording way: Real-time Recording, Review Printing, Periodic Recording, Alarm Recording
3.6 Alarm Details
- Level: Low, medium and high
- Indication: Auditory and visual
- Alarm volume: Adjustable
- Alarm pause time: 1min, 2min
- Parameter alarm type: Latch/ Unlatch
3.7 Input Device
- Knob (Standard)
- Keypad input (Standard)
- Mouse/ Keyboard input (option)
3.8 System Output & Extensible Interface
- Ethernet Network: 1 standard RJ45 socket
- USB Port: 1
- Video Output: 1 HDMI port (option)
3.9 Operating Environment
- Temperature: 5 ~ 40 ℃
- Humidity: 15% ~ 80% (non-condensing)
- Atmosphere pressure: 86 KPa ~ 110 KPa
3.10 Transportation and Storage Environment
- Temperature: -20~50 ℃
- Humidity: 10%~90% (non-condensing)
- Atmosphere pressure: 86 KPa ~ 110 KPa
3.11 Safety
- IEC60601-1 Approved, CE marking according to MDD93/42/EEC
- With reference to RoHS Directive 2011/65/EU recasting
3.12 Trend & Reviewing
- Trend: 2160 hours
- ARR events: 128 groups of ARR events and associated waveform
- NIBP measurement reviewing: 2000 groups
- Waveform review: 2 hours
- Alarm event: 2000 groups of parameter alarms events and associated parameter
4. Monitoring Parameters Specifications
4.1 ECG
- Lead mode: 3/5 Leads (I, II, III or I, II, III, AVR, AVL, AVF, V)
- Protection: Breakdown Voltage 4000VAC 50/60Hz; Defibrillator proof
- Gain: 2.5mm/mV (×0.25), 5.0mm/mV (×0.5), 10mm/mV (×1), 20mm/mV (×2), 40 mm/ mV (×4), Auto
- Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
- ECG signal range: ±12 mV (Gain × 0.25)
- Accuracy: ±1bpm/ ±1%, whichever is greater
- Resolution: 1 bpm
- Leakage Current: < 10 μA
- Baseline Recovery: ≤3s after defibrillation (Monitor mode); ≤1s after defibrillation (Surgery mode)
- Bandwidth: Surgery 1 ~ 25 Hz; Monitor 0.5 ~ 40 Hz; Diagnostic 0.05 ~ 120 Hz
- Indication of Electrode Separation: Every electrode (exclusive of RL)
4.2 Heart Rate
- Measure range: Adult: 0bpm, 10 ~ 300 bpm; Neo/Ped: 0bpm, 10 ~ 350 bpm
- Resolution: 1 bpm
- Accuracy: ± 1% or ± 1bpm, whichever is greater
4.3 ST Measurement
- Range: -2.0 ~ +2.0 mV
- Accuracy: -0.8mV ~+0.8mV: ±0.02mV or ±10%, whichever is greater; Other range: unspecified
- Resolution: 0.01mV
4.4 Respiration
- Method: Impedance between RA-LL, RA-LA
- Gain: ×0.25, ×0.50, ×1, ×2, ×4
- Respiration Rate: 0bpm, 6 ~ 150 BrPM
- Sweep speed: 6.25 mm/s, 12.5 mm/s, 25mm/s
- Resolution: 1 BrPM
- Accuracy: ±2 bpm or ±2%, whichever is greater
- Apnea Alarm: 10 ~ 60 s
4.5 NIBP
- Method: Oscillometric
- Measure mode: Manual, Auto, STAT
- Measure Interval in AUTO Mode: 1~480 min
- STAT mode cycle time: Keep 5 minutes, at 5 seconds interval
- Measure and Alarm Range:
Group Parameter Range Adult SYS: 30 ~ 270 mmHg; DIA: 10 ~ 220 mmHg; MEAN: 20 ~ 235 mmHg Pediatric SYS: 40 ~ 235 mmHg; DIA: 10 ~ 220 mmHg; MEAN: 20 ~ 225 mmHg Neonate SYS: 30 ~ 135 mmHg; DIA: 10 ~ 110 mmHg; MEAN: 20 ~ 125 mmHg - Static pressure accuracy: ±3mmHg
- Resolution: 1mmHg
- Accuracy: Maximum Mean error ±5mmHg; Maximum Standard deviation ≤8mmHg
- Over pressure Protection: Dual protection via software & hardware
4.6 Temperature
- Technique: Thermistor probe (2.25K)
- Channel: Dual-Channel, provide T1; T2; ΔT
- Measuring and Alarm Range: 0.0 ℃~ 50 ℃ (32℉~ 122℉)
- Unit: Celsius (℃), Fahrenheit (℉)
- Resolution: 0.1℃ or 1℉
- Accuracy: ±0.1℃ (exclusive probe)
4.7 SpO2 (Digital Technic)
- Measurement Range: 0 ~ 100 %
- Resolution: 1 %
- Response Modes: Low, Medium, High
- Accuracy: ±2% (70% ~ 100%); ±3 % (35% ~ 69%); Unspecified (0 ~ 34%)
- Support Pitch tone and multi-level volume
4.8 Pulse Rate
- Measuring and Alarm Range: 25~250bpm
- Accuracy: ±1% or ±1 bpm, whichever is greater
- Resolution: 1bpm
4.9 Nellcor-SpO2 (Option)
- Measurement Range: 0 ~ 100 %
- Resolution: 1 %
- Accuracy: 70% ~ 100%, ±2 % (adult); 70% ~ 100%, ±3 % (Neonate); 70% ~ 100%, ±2 % (Low Perfusion); 0% ~ 69%, unspecified
- Pulse Rate (Nellcor-SpO2):
- Measurement range: 20 ~ 300 bpm
- Resolution: 1bpm
- Accuracy: ±3 bpm (20 ~ 250 bpm); unspecified (251~300 bpm)
4.10 Masimo SpO2 (Option)
- Measurement Range: 0 ~ 100 %
- Resolution: 1 %
- Accuracy: 70% ~ 100%, ±2 % (adult/ pediatric, non-motion conditions); 70% ~ 100%, ±3 % (neonate, non-motion conditions); 70% ~ 100%, ±3 % (motion conditions); 0% ~ 69% unspecified
- Pulse Rate (Masimo SpO2):
- Measurement range: 25 ~ 240 bpm
- Resolution: 1bpm
- Accuracy: ±3 bpm (non-motion condition)
4.11 EtCO2 (Mainstream/ Sidestream) (Option)
- Measure method: Non-dispersive infrared (NDIR)
- Measure Range: 0 ~19.7% (0 ~ 150 mmHg), 0 ~ 20 kPa
- Resolution: 0.1 mmHg
- CO2 Accuracy: 0 ~ 40 mmHg, ±2 mmHg; 41 ~ 70 mmHg, ±5% of reading; 71 ~ 100 mmHg, ±8% of reading; 101~ 150 mmHg, ±10% of reading (at 760 mmHg, ambient temperature of 25℃)
- Respiratory Rate: Range: 3 ~150 BrPM; Accuracy: ±1 bpm
4.12 EtCO2 (Micro-stream) (Option)
- Measure method: Non-dispersive infrared (NDIR)
- Measure Range: 0 ~19.7% (0 ~150 mmHg), 0 ~ 20 kPa
- Sample Rate: 50 mL/min ±10mL/min
- Resolution: 0.1 mmHg (0 ~ 50 mmHg); 0.25 mmHg (50 ~ 114 mmHg)
- CO2 Accuracy: 0 ~ 40 mmHg, ±2 mmHg; 41 ~ 70 mmHg, ±5% of reading; 71 ~ 100 mmHg, ±8% of reading; 101~ 150 mmHg, ±10% of reading (at 760 mmHg, ambient temperature of 35℃)
- Respiratory Rate: Range: 3 ~120 BrPM; Accuracy: ±1 bpm
4.13 Depth of Anesthesia (CSI) (Option)
- EEG sensitivity: ± 400μV
- Noise: < 2μVp-p, < 0.4μV RMS, 1-250 Hz
- CMRR: > 140 dB
- Input impedance: > 50 Mohm
- Sample rate: 2000 samples/sec, (14 bits equivalent)
- BS%: 0-100, filter 1-42 Hz, 1 sec. display update
- EMG: 0-100 Logarithmic. Filter 75-85 Hz, 1 sec. update
- Alarms: High / Low with user selectable limit
- Artifact rejection: Automatic
- Sensor impedance range: 0 – 10 kOhm / measurement current 0.01 μA
4.14 Multi-gas/O2 (Anesthetic Gas) (Option)
- Method: Infrared absorption
- Gas sorts: CO2, N2O, Des, Iso, Enf, Sevo, Hal, O2 (Optional paramagnetic sensor)
- Calibration: Room air calibration performed automatically when changing airway Airway adapt (< 5 sec)
- Measurement range: CO2: 0~25%; N2O: 0~100%; O2: 0~100%; Enf, Iso, Hal: 0~25%; Sevo, Des: 0~25%
- Data output: Fi and ET values
- Respiration rate: 0~ 150 BrPM
- Other: Up to 5 waveforms displayed; Agent mixture detection; MAC value displayed
4.15 IBP (Option)
- Max Channel: 2
- Measurement way: Thermal resistance way
- Press Sensor: Sensitivity 5 uV/V/mmHg, ±2%; Impedance 300 to 3000Ω
- Resolution: 1 mmHg
- Unit: mmKg, kPa, cmH2O
- Transducer sites: Arterial Pressure (ART); Pulmonary Arterial (PA); Left Arterial (LAP); Right Arterial (RAP); Central Venous Pressure (CVP); Intracranial Pressure (ICP); P1/ P2
- Measuring and alarm range:
- ART: 0 ~ +350mmHg
- PA: -10 ~ +120 mmHg
- CVP/ RAP/ LAP/ ICP: -10 ~ +40 mmHg
- P1/ P2: -50 ~ +350mmHg
- Accuracy:
- Static: ±1mmHg or ±2%, whichever is greater (exclusive of transducer); ±4mmHg or ±4%, whichever is greater (inclusive of transducer)
- Dynamic: ±4mmHg or ±4%, whichever is greater
5. Model Configuration
| Type | Configuration Details |
|---|---|
| Standard Config. | 15 inch LED, 3/5 Lead ECG, NIBP, SpO2, Pulse Rate, Temperature, Respiration |
| Optional Config. | 1/2 IBP, 2-Temperature, Nellcor SpO2, Masimo SpO2, 12-Lead ECG, EtCO2 (Mainstream/Sidestream/Microstream), Anesthesia Depth Monitoring (CSM module), Multi-gas (AG) Monitoring |
| Optional Accessories | Printer, USB Mouse/Computer Keyboard Input Function, Central Monitoring Station, Neo/Ped Accessories, HDMI Output, Wall-mounting, Trolley |
| Remark | User can NOT choose USB mouse function and WIFI at the same time |